产品参数(Product Specifications)
Assay TypeSandwich-ELISAAnalyteIL-10Format96T(8×12 strips)ReactivityHumanRegulatory StatusRUOSensitivity<6.25 pg/mLStandard Curve Range6.25 pg/mL-200 pg/mLAssay Time2 hr 50 minSuitable Sample TypeFor detection and quantitative determination of GMP human IL-10 in samples from CAR-T product preparation processing.Sample volume100 uL产品概述(Product Overview)
resDetect™ Human Interleukin-10 (IL-10) ELISA Kit (Residue Testing) is based on the ELISA sandwich method and is used to detect and quantitatively determine GMP human IL-10 in cell culture supernatants, serum, and plasma. The kit contains GMP human IL-10 (ACROBiosystems, cat#GMP-L10H25) to ensure accurate assay results. It is designed to provide a reliable solution for CAR-T product quality assessment during drug development and CMC quality control stages. Additionally, it can be used as a universal detection tool for the quantitative determination of human IL-10.
重构方法(Reconstitution)
Please see Certificate of Analysis for details of reconstitution instruction and specific concentration.
存储(Storage)
1. Unopened kit should be stored at 2℃-8℃ upon receiving.
2. Find the expiration date on the outside packaging and do not use reagents past their expiration date.
3. The opened kit should be stored per components table. The shelf life is 30 days from the date of opening.
组分(Materials Provided)
IDComponentsSizeCRS008-C01Pre-coated Anti-IL-10 Antibody Microplate1 plate(8×12 strips)CRS008-C02Human IL-10 Standard20 μgCRS008-C03Biotin-Anti-IL-10 Antibody50 μLCRS008-C04Streptavidin-HRP50 μLCRS008-C0510×Washing Buffer50 mLCRS008-C062×Dilution Buffer50 mLCRS008-C07Substrate Solution12 mLCRS008-C08Stop Solution7 mL质量管理控制体系(QMS)
数据展示
典型数据-Typical Data
Please refer to DS document for the assay protocol.
For each experiment, a standard curve needs to be set for each micro-plate, and the specific OD value may vary depending on different laboratories, testers, or equipments. The following example data is for reference only.
稀释线性(Dilution Linearity)
To assess the linearity of the assay, samples spiked with high concentrations of human IL-10 were serially diluted with calibrator diluent to produce samples with values within the dynamic range of the assay.
批内差异(Intra-Assay Statistics)
Three samples of known concentration were tested ten times on one plate to assess intra-assay precision.
批间差异(Inter-Assay Statistics)
Three samples of known concentration were tested in three separate assays to assess inter-assay precision.
回收率(Recovery)
Five parts of blank T cell culture supernatant were added with different concentrations of human IL-10, and the T cell culture supernatant without human IL-10 was used as background to calculate the recovery rate. The range of the recovery rate is 86.6-95.0%, and the average recovery is 91.9%.